Specifications

MODEMETHOD OF MEASUREMENT
HoldDisplays real-time readings until stable or 3 minutes, whichever comes first, then displays final readings.
LIVEDisplays real-time readings continuously for up to 5 minutes.
MODEPARAMETER MEASURED
DO DefaultDisplay alternates between DO Concentration in ppm, DO Saturation in % and Temperature.
DO ConcentrationDisplay alternates between DO Concentration in ppm and Temperature.
DO SaturationDisplay alternates between DO Saturation in % and Temperature.
SpecificationsDO ConcentrationDO SaturationTemperature
Ranges0-20 ppm0-200%0 – 71°C / 32 – 160°F
Accuracy1± 0.2 ppm or ± 2% of the reading, whichever is greater (up to 40ºC)0.0% to ≤ 100.0%: ± 20 counts; >100%: ± 2% of the reading.± 0.1°C
Resolution0.01 ppm0.1%0.1°C/F
Altitude Adjustment:-2,000 ft. to 15,000 ft. in increments of 100 ft.
Salinity Adjustment:0 – 50 ppt in increments of 1 ppt
Time to Reading Stabilization:10 – 180 seconds
Calibration Methods:1 or 2 Point; 0 ppm DO Calibration Solution, Air or Water
Operating/Storage Temperature:0 – 55°C / 32 – 131°F
Power Consumption:Active Mode: 30 – 140 mA, Sleep Mode: 2 µA
Temperature Compensation for Membrane Permeability:Automatic
Physical Dimensions:17.15 cm L x 1.59 cm D or 6.75 in. L x .625 in. D
Weight:54 g or 1.94 oz.
Case Material:Anodized Aircraft Aluminum with Protective Coating
Battery Type:N type, Alkaline 1.5 V
Water Resistance:IP67 and NEMA 6
EN61236-1: 2006 – Annex A: 2008; 1999/5/EC Annex IIIElectrostatic discharge to case of instrument may cause PT5 to spontaneously power ON. In this case, the PT5 will power OFF after several seconds.